Detail Cantuman

Image of INFORMASI CITRA ANATOMI PADA PENGGUNAAN VARIASI INCREMENT PEMERIKSAAN MSCT ABDOMEN IRISAN AXIAL KASUS NODUL HEPAR

Skripsi D IV

INFORMASI CITRA ANATOMI PADA PENGGUNAAN VARIASI INCREMENT PEMERIKSAAN MSCT ABDOMEN IRISAN AXIAL KASUS NODUL HEPAR

XML

Liver nodules are tumors that are less than 2 cm in size and are single or multiple. In making a diagnosis of liver nodules can be done by examination of
abdominal MSCT. According to Seeram (2009), using 50% overlapping there was an increase in nodule detection by 10% compared to incremental increments. The purpose of the study was to determine differences in information on anatomical images on the use of increment variations in abdominal MSCT examination and to find the appropriate increment values to produce optimal anatomical image information.
This type of research is quantitative with an experimental approach. The study was conducted at the RSUD. Tugurejo Semarang, the image was taken from abdominal MSCT examination in 5 patients with a variation of increments of 100% (8 mm), 50% (4 mm), 30% (2, 4 mm). The results of the study were assessed by 3 respondents. Data analysis used the kappa statistical test, crosstabulation, and friedman.
From the statistical test, the results of the 2.4 mm increment value show clear anatomical criteria and can detect small nodules. Supported by using friedman statistical test p-value = 0.000 which means that p-value


Informasi Detail

Pernyataan Tanggungjawab
dr. Arinawati, Sp. Rad dan Darmini, S.Si, M.Kes
Pengarang
SELFIA CLAUDIANA RAHMA DEWI - Pengarang Utama
NIM
P1337430215055
Bahasa
Indonesia
Deskripsi Fisik
xvi + 62 p.; Bibl.; Ilus.; 21 cm x 29 cm
Dilihat sebanyak
723
Penerbit Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG : Semarang.,
Edisi
Subjek
Klasifikasi

Lampiran Berkas

Citation
SELFIA CLAUDIANA RAHMA DEWI. (2019).INFORMASI CITRA ANATOMI PADA PENGGUNAAN VARIASI INCREMENT PEMERIKSAAN MSCT ABDOMEN IRISAN AXIAL KASUS NODUL HEPAR().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG

SELFIA CLAUDIANA RAHMA DEWI.INFORMASI CITRA ANATOMI PADA PENGGUNAAN VARIASI INCREMENT PEMERIKSAAN MSCT ABDOMEN IRISAN AXIAL KASUS NODUL HEPAR().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2019.Skripsi D IV

SELFIA CLAUDIANA RAHMA DEWI.INFORMASI CITRA ANATOMI PADA PENGGUNAAN VARIASI INCREMENT PEMERIKSAAN MSCT ABDOMEN IRISAN AXIAL KASUS NODUL HEPAR().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2019.Skripsi D IV

SELFIA CLAUDIANA RAHMA DEWI.INFORMASI CITRA ANATOMI PADA PENGGUNAAN VARIASI INCREMENT PEMERIKSAAN MSCT ABDOMEN IRISAN AXIAL KASUS NODUL HEPAR().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2019.Skripsi D IV



Dirujuk oleh 0 dokumen