Backgroud: The suggested parameter protocol on the MRI Cervical one of the GRE (T2*W) with a thin slice that is angled so that it can align with the disk space or perpendicular to the lesion in the examination. Recommended pulse sequences on MRI Cervical on axial using Flip angle = 200 and Flip angle = 300. Previous studies have found results Flip angle of 100 or 150 provide better image contrast in the spinal cord and subarachnoid space around it than the flip angle 300. In Mayapada Hospital South Jakarta, MRI Cervical on axial, Radiographer using the flip angle 250. The aim of this research is to know the difference of flip angle variation and to get the best flip angle angle for MRI Cervical.
Methods: The type of this research is quantitative research with experimental approach. The sample used in this study were 10 volunteers who were examined by MRI 3 tesla at Mayapada Hospital South Jakarta. Each sample was examined with flip angle variations of 50, 100, 200, and 250 on the MRI Cervical of the axial slice. Image results were assessed by 2 radiologist specialists. Then, Image results are tested using friedman test.
Results: The friedman test of variation of flip angle 50, 100, 200, and 250 got the result 0,004 from significance ρ value


Informasi Detail

Pernyataan Tanggungjawab
Sudiyono; Dwi Rochmawat
Pengarang
Muhammad Wildan Hayadin - Pengarang Utama
NIM
P1337430217114
Bahasa
Indonesia
Deskripsi Fisik
xii + 64 HLM.: ILLUS.: LAMP. : TABEL ; 29
Dilihat sebanyak
989
Penerbit Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG : Semarang.,
Edisi
Subjek
Klasifikasi
081

Lampiran Berkas

Citation
Muhammad Wildan Hayadin. (2018).VARIASI FLIP ANGLE SEKUENS GRADENT ECHO (T2*W) TERHADAP INFORMASI ANATOMI PADA PEMERIKSAAN MRI VERTEBRA SERVIKALIS POTONGAN AXIAL().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG

Muhammad Wildan Hayadin.VARIASI FLIP ANGLE SEKUENS GRADENT ECHO (T2*W) TERHADAP INFORMASI ANATOMI PADA PEMERIKSAAN MRI VERTEBRA SERVIKALIS POTONGAN AXIAL().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2018.Skripsi D IV

Muhammad Wildan Hayadin.VARIASI FLIP ANGLE SEKUENS GRADENT ECHO (T2*W) TERHADAP INFORMASI ANATOMI PADA PEMERIKSAAN MRI VERTEBRA SERVIKALIS POTONGAN AXIAL().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2018.Skripsi D IV

Muhammad Wildan Hayadin.VARIASI FLIP ANGLE SEKUENS GRADENT ECHO (T2*W) TERHADAP INFORMASI ANATOMI PADA PEMERIKSAAN MRI VERTEBRA SERVIKALIS POTONGAN AXIAL().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2018.Skripsi D IV



Dirujuk oleh 0 dokumen