Skripsi D IV
ANALISA PERBEDAAN CNR DAN INFORMASI ANATOMI MRI CERVICAL PADA SEKUEN T2 TSE POTONGAN SAGITAL DENGAN VARIASI ECHO TRAIN LENGTH (ETL)
XMLBackgground : In Semarang Telogorejo Hospital and Surabaya Haji Hospital used different ETL values on sequences T2 TSE of MRI Cervical examination ie ETL 22 and ETL 18, in theory (Westbrook and Kaunt, 2014) the ETL value for T2 is 16+. Because in theory difference of ETL value will cause different image quality also The purpose of this study was to determine whether or not ETL values in ETL’s 18, 20 and 22 were to obtain optimal ETL values and resulted in the best quality on T2 TSE sagittal MRI Cervical sequences on MRI 1.5 T.
Methods : This research is an experimental quantitative research. This research was conducted in SMC RS Telogorejo Semarang using MRI Siemens 1.5 T. Data were taken from 10 probandus with 3 variations of ETL (18,20 and 22). Image assessment was done quantitatively with Region Of Interest (ROI) method on CNR and with assessment subjective of two radiologist doctor on anatomical information according to the level of clarity in corpus, discuss, spinal cord, cerebro spinal fluid. The data were analyzed by friedman statistic test, then wilcoxon statistic test on CNR and anatomy information.
Results : The results showed that there was a difference of CNR value with p-value 0.045 and anatomical information on ETL variation with p-value
Informasi Detail
Pernyataan Tanggungjawab |
RINI INDRATI, S.Si, M.Kes DAN EDY SUSANTO, SH, S.Si M.Kes
|
---|---|
Pengarang |
MUHAMMAD HARI SAFANTRI - Pengarang Utama
|
NIM |
P1337430214036
|
Bahasa |
Indonesia
|
Deskripsi Fisik |
xv+ 76 hlm.'; ilus.; tab.; 29 cm
|
Dilihat sebanyak |
742
|
Penerbit | Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG : POLTEKKES KEMENKES SEMARANG., 2018 |
---|---|
Edisi | |
Subjek | |
Klasifikasi |
081
|