Skripsi D IV
ANALISIS PERBEDAAN INFORMASI ANATOMI DAN KUALITAS CITRA ANTARA PENGGUNAAN HEAD COIL DENGAN FLEX M COIL PEMERIKSAAN MRI ANKLE JOINT PADA PROTON DENSITY WEIGHTED SPAIR
XMLBackground : Examination at MRI such as MRI ankle joint examination have to use radiofrequency coil for transmitter and receiver signal radiofrequency to product better SNR and CNR. At some hospital ankle joint used head coil for specific case, but at the literature using flexible coil. The purpose of this study was to determine the difference of anatomical information and image quality to define the better quality image ankle joint and coil radiofrequency that are the best for ankle joint examination.
Methods :This research type is experiment. This research was conducted at Radiology Installation of Siloam Hospitals Surabaya with sample of 8 MRI ankle joint volunteer. Image assessment was done qualitatively and in the form of assessment from two respondents about the clarity of anatomical information including talocrural joint, os. talus, and Achilles tendon. Qualitative data analysis was conducted by Cohen's Kappa test to test the reliability level among the two respondents, Wilcoxon test and paired t-test.
Results : The result showed that there was significant differences between using flex m coil and head coil that showed by p value/sig : 0,002 for anatomy information and for image quality p value/sig : 0,000 that means p value/sig
Informasi Detail
Pernyataan Tanggungjawab |
HERMINA SUKMANINGTYAS DAN BAGUS ABIMANYU
|
---|---|
Pengarang |
YUNITA AJENG PRASDIYANTI - Pengarang Utama
HERMINA SUKMANINGTYAS - Pengarang Tambahan BAGUS ABIMANYU - Pengarang Tambahan |
NIM |
P17430213077
|
Bahasa |
Indonesia
|
Deskripsi Fisik |
XVI, 85 pages:28 cm
|
Dilihat sebanyak |
1513
|
Penerbit | DIV TEKNIK RADIOLOG : Semarang., 2017 |
---|---|
Edisi | |
Subjek | |
Klasifikasi |
081
|
Lampiran Berkas
Dirujuk oleh 1 dokumen
- (2019) Chanafi Setyana Putra