Detail Cantuman

Image of STANDARISASI INDEKS EKSPOSUR UNTUK MEMENUHI KRITERIA ANATOMI DAN ASPEK TEKNIS PADA THORAX PEDIATRIK DI RSUD DR. LOEKMONO HADI KUDUS

Tugas Akhir DIII

STANDARISASI INDEKS EKSPOSUR UNTUK MEMENUHI KRITERIA ANATOMI DAN ASPEK TEKNIS PADA THORAX PEDIATRIK DI RSUD DR. LOEKMONO HADI KUDUS

XML

In CR there is an exposure index that used as an exposure indicator. But in Kudus General Hospital, the resulting of exposure index value was ignored, and in the preliminary survey of pediatric chest examination resulting a large exposure index range. Research has been done in Kudus General Hospital about standardization of exposure index on pediatric chest examination. The aim of this study is to know the profile of exposure index value and the setting of the exposure factors, the assessment of anatomy criteria and technical aspect, and the right exposure factors such as kV and mAs on pediatric chest examination in Kudus General Hospital.
The type of this research is descriptive quantitative. The research was done by collecting data related to pediatric chest radiograph (≤ 2 years) from March – April 2018 and the value of exposure index was recorded, then the radiograph was assessed using questionnaires that filled by one respondent who is a radiologist. The data was analyzed by displaying the data of exposure index and anatomy criteria from questionnaires into the table form, and then conducted a descriptive analysis to be drawn conclusions.
The results showed the profile of exposure index value and the setting of the exposure factor in Kudus General Hospital has a fairly large exposure index range of 1084 – 2175, using 40 kV and 10 mAs and the collimation still often exceeds the object. Then for the assessment of the thorax anatomical criteria and the technical aspect overall was “Good Enough”, and the right exposure factors, that is: at 6 and 7 cm chest thickness was using 60 kV and 1,6 mAs; at 11 cm chest thickness was using 60 kV and 2 mAs, FFD 100 cm, and the setting of collimation as wide as object, the exposure index generated in the normal range that is 1251 – 1382.


Informasi Detail

Pernyataan Tanggungjawab
Dwi Rochmayanti, S.ST., M.Eng.
Pengarang
Mukhammad Lutfan Nurrokhim - Pengarang Utama
NIM
P1337430115007
Bahasa
Indonesia
Deskripsi Fisik
xii + 59 hlm,; Bibl.;Ilus.; 29 cm
Dilihat sebanyak
854
Penerbit Prodi DIII T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG : Semarang.,
Edisi
Subjek
Klasifikasi
081

Lampiran Berkas

Citation
Mukhammad Lutfan Nurrokhim. (2018).STANDARISASI INDEKS EKSPOSUR UNTUK MEMENUHI KRITERIA ANATOMI DAN ASPEK TEKNIS PADA THORAX PEDIATRIK DI RSUD DR. LOEKMONO HADI KUDUS().Semarang:Prodi DIII T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG

Mukhammad Lutfan Nurrokhim.STANDARISASI INDEKS EKSPOSUR UNTUK MEMENUHI KRITERIA ANATOMI DAN ASPEK TEKNIS PADA THORAX PEDIATRIK DI RSUD DR. LOEKMONO HADI KUDUS().Semarang:Prodi DIII T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2018.Tugas Akhir DIII

Mukhammad Lutfan Nurrokhim.STANDARISASI INDEKS EKSPOSUR UNTUK MEMENUHI KRITERIA ANATOMI DAN ASPEK TEKNIS PADA THORAX PEDIATRIK DI RSUD DR. LOEKMONO HADI KUDUS().Semarang:Prodi DIII T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2018.Tugas Akhir DIII

Mukhammad Lutfan Nurrokhim.STANDARISASI INDEKS EKSPOSUR UNTUK MEMENUHI KRITERIA ANATOMI DAN ASPEK TEKNIS PADA THORAX PEDIATRIK DI RSUD DR. LOEKMONO HADI KUDUS().Semarang:Prodi DIII T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2018.Tugas Akhir DIII



Dirujuk oleh 0 dokumen