Image of PENGUKURAN LAJU PAPARAN RADIASI PADA RUANG CT-SCAN DI INSTALASI RADIOLOGI RSUD Ir. SOEKARNO KABUPATEN SUKOHARJO

Tugas Akhir DIII

PENGUKURAN LAJU PAPARAN RADIASI PADA RUANG CT-SCAN DI INSTALASI RADIOLOGI RSUD Ir. SOEKARNO KABUPATEN SUKOHARJO

XML

CT-Scan is an imaging modality in the field of diagnostics, especially radiology that produces large radiation, so should pay attention to aspects of protection and radiation safety for radiographers and public. At the CT-Scan examination room at Radiology Installation of RSUD Ir. Soekarno has made door replacement and addition of 2 mm Pb on the window and has not done measurement of exposure rate of re-radiation. This study aims to determine the value of the radiation exposure rate and the safety of the examination room on radiation exposure for workers and public reffering to the publication of NCRP no. 147 on the design of protective structures for health facilities.
This type of research is quantitative with survey approach. This research was conducted in May 2018. Data collection was done by observation, documentation, measurement of exposure rate at a number of points. Data analysis was done descriptively by comparing data with NCRP report no 147 for controlled areas is 2,5 µSv/h and uncontrolled area is 0,5 µSv/h.
The radiation exposure rate measurement results showed the presence of excessive radiation exposure in controlled areas, namely the Pb glass in the operator's room of 29.5 μSv/h and the uncontrolled area at the patient's entrance point was 1.75 μSv/h. This value exceeds the NCRP recommendation value no. 147 that is in a controlled area of 2.5 μSv/h and uncontrolled area of 0.5 μSv/h. Radiology Installation of RSUD Ir. Soekarno Sukoharjo Regency is located at an altitude of 103 meters above sea level background radiation exposure value of 1.225 μSv/h, which exceeds the provisions of the background exposure rate according to UNSCEAR for a height of 0 meters above sea level is 0.03 μSv/h Leakage of radiation exposure that occurs at the patient's entrance and Pb glass in the operator's station is not safe for radiographers and public because it exceeds the permissible radiation exposure rate limit and it is recommended to replace Pb glass in the operator's room with a thickness of 2 mm Pb.


Informasi Detail

Pernyataan Tanggungjawab
Sudiyono, SE, M.Kes
Pengarang
DEVI FEBRIANA PUTRI - Pengarang Utama
NIM
P1337430115005
Bahasa
English
Deskripsi Fisik
xv; 70 hal.; ilus.; 29 cm
Dilihat sebanyak
1337
Penerbit Poltekkes Kemenkes Smarang : Semarang.,
Edisi
Subjek
Klasifikasi
081

Lampiran Berkas

Citation
DEVI FEBRIANA PUTRI. (2018).PENGUKURAN LAJU PAPARAN RADIASI PADA RUANG CT-SCAN DI INSTALASI RADIOLOGI RSUD Ir. SOEKARNO KABUPATEN SUKOHARJO().Semarang:Poltekkes Kemenkes Smarang

DEVI FEBRIANA PUTRI.PENGUKURAN LAJU PAPARAN RADIASI PADA RUANG CT-SCAN DI INSTALASI RADIOLOGI RSUD Ir. SOEKARNO KABUPATEN SUKOHARJO().Semarang:Poltekkes Kemenkes Smarang,2018.Tugas Akhir DIII

DEVI FEBRIANA PUTRI.PENGUKURAN LAJU PAPARAN RADIASI PADA RUANG CT-SCAN DI INSTALASI RADIOLOGI RSUD Ir. SOEKARNO KABUPATEN SUKOHARJO().Semarang:Poltekkes Kemenkes Smarang,2018.Tugas Akhir DIII

DEVI FEBRIANA PUTRI.PENGUKURAN LAJU PAPARAN RADIASI PADA RUANG CT-SCAN DI INSTALASI RADIOLOGI RSUD Ir. SOEKARNO KABUPATEN SUKOHARJO().Semarang:Poltekkes Kemenkes Smarang,2018.Tugas Akhir DIII



Dirujuk oleh 0 dokumen