3D T2-VISTA is modification sequences and FSE is the sequential sequences of Spin Echo. 3D T2-VISTA is a sequence that can produce data from sagital, axial and coronal slices with the same time as FSE sequence. The aim of this study was to determine the diagnostic performance of 3D T2-VISTA with FSE in describing anatomy, image quality also sensitivity and specificity in assessing Hernia Nukleus Pulposus cases.
This research type is quantitative with an experimental approach. This study was conducted at Cempaka Putih Hospital in Jakarta using 10 patients with Hernia Nukleus Pulposus cases. Image results of FSE and 3D T2-VISTA were reviewed by 3 radiologists. Image quality was assessed by giving ROI to anatomical structures. The next study was to assess sensitivity and specificity. Data analysis was performed by cohen’s kappa to see the agreement level among respondents followed by the Wilcoxon test to assess anatomical differences and image quality between FSE and 3D T2-VISTA. Futhermore, assessing sensitivity and specificity to determine diagnostic performance value .
The test results stated that there were differences in anatomic information with p-Value 0,05 and show better visualization anatomic structure in 3D T2-VISTA than FSE.. SNR and CNR values are higher in 3D T2-VISTA than FSE. The diagnostic performance showed the difference with the sensitivity (83.3%) and specificity (75%) in 3D T2-VISTA higher than T2-FSE sensitivity (66.67%) and specificity (50%). So, 3D T2-VISTA is superior in describing anatomy because it has a high resolution, short echo spacing and reduces artifact pulsation, also assesses Hernia Nukleus Pulposus better


Informasi Detail

Pernyataan Tanggungjawab
KARINA WIDYA NASTITI
Pengarang
KARINA WIDYA NASTITI - Pengarang Utama
Bekti Safarini - First Advisor
Sri Mulyati S - Second Advisor
NIM
P1337430215066
Bahasa
Indonesia
Deskripsi Fisik
xvii + 107 hlm,;Bibl.;lus.;29 cm
Dilihat sebanyak
1247
Penerbit DIV TEKNIK RADIOLOG : Semarang.,
Edisi
Subjek
Klasifikasi

Lampiran Berkas

Citation
KARINA WIDYA NASTITI. (2019).PERFORMA DIAGNOSTIK MRI CERVICAL SEKUEN 3D T2-VISTA DENGAN T2-FSE POTONGAN SAGITAL PADA KASUS HERNIA NUKLEUS PULPOSUS().Semarang:DIV TEKNIK RADIOLOG

KARINA WIDYA NASTITI.PERFORMA DIAGNOSTIK MRI CERVICAL SEKUEN 3D T2-VISTA DENGAN T2-FSE POTONGAN SAGITAL PADA KASUS HERNIA NUKLEUS PULPOSUS().Semarang:DIV TEKNIK RADIOLOG,2019.Skripsi D IV

KARINA WIDYA NASTITI.PERFORMA DIAGNOSTIK MRI CERVICAL SEKUEN 3D T2-VISTA DENGAN T2-FSE POTONGAN SAGITAL PADA KASUS HERNIA NUKLEUS PULPOSUS().Semarang:DIV TEKNIK RADIOLOG,2019.Skripsi D IV

KARINA WIDYA NASTITI.PERFORMA DIAGNOSTIK MRI CERVICAL SEKUEN 3D T2-VISTA DENGAN T2-FSE POTONGAN SAGITAL PADA KASUS HERNIA NUKLEUS PULPOSUS().Semarang:DIV TEKNIK RADIOLOG,2019.Skripsi D IV



Dirujuk oleh 0 dokumen