Skripsi D IV
PENGARUH VARIASI TIME REPETITION (TR) TERHADAP KUALITAS CITRA DAN INFORMASI CITRA PADA PEMERIKSAAN MRI LUMBAL SEKUENS T2 FSE POTONGAN SAGITAL
XMLTR (Time Repetition) is an MRI parameter. TR controls the amount of longitudinal magnetization recovery before the next RF pulse. TR parameter is one of parameters that can affect image quality and image information. Long TR arrangements can increase SNR (Signal To Noise Ratio), CNR (Contrast To Noise Ratio) and provide better contrast, but increase the examination time. The purpose of this research is to know the effect of TR to image quality, difference of image information and TR optimal on MRI Lumbal sequence T2 FSE sagittal.
The type of this research is quantitative research with experimental approach. The sample of research used is 6 probandus. TR variations used are TR 2000, 2500, 3000, 3500, 4000 with other parameters controlled. Aspects of image quality and image information values assessed in this study include Corpus Vertebrae, Discus Vertebrae, Cerebro Spinal Fluid (CSF), Medulla Spinalis, and Soft Tissue. Processing and analysis of image quality data is done by data normality test using shapiro-wilk, then test influence correlation pearson, linear regression and repeated anova. and for image information is done normality data test using shapiro-wilk, and friedman test.
The results of the test analysis showed that there was a TR effect on SNR image quality (p = 0.005) and CNR (p = 0.043) with (p
Informasi Detail
Pernyataan Tanggungjawab |
Fatimah, S.ST, M.Kes ; M. Irwan Katili, S.Pd, M.Kes
|
---|---|
Pengarang |
SYAFIUDDIN NIZAR - Pengarang Utama
|
NIM |
P1337430214017
|
Bahasa |
Indonesia
|
Deskripsi Fisik |
xv, 90 hlm.:ilus:tabel:lampiran ;29 cm
|
Dilihat sebanyak |
1148
|
Penerbit | Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG : POLTEKKES KEMENKES SEMARANG., 2018 |
---|---|
Edisi |
2018
|
Subjek | |
Klasifikasi |
081
|