Image of PERBEDAAN INFORMASI CITRA ANATOMI PADA PEMERIKSAAN MSCTNASOFARING POTONGAN AXIAL DENGAN MENGGUNAKAN VARIASI KERNEL

Skripsi D IV

PERBEDAAN INFORMASI CITRA ANATOMI PADA PEMERIKSAAN MSCTNASOFARING POTONGAN AXIAL DENGAN MENGGUNAKAN VARIASI KERNEL

XML

Algorithhm reconstruction is a mathematical procedure used in recontructing images.The characterustics of the CT-Scan image depends on the strength of the selected algorithm. The higher the resolution of the selected algorithm the higher the resolution of the image will be generated. In radiology installation RSUD Tugurejo Semarang on the examination of nasopharyngeal CT-Scan using kernel head parameters.this study aims to determine the difference of anatomical image information and to determine the most optimal kernel value.`
This type of research is quantitative research with eksperimental approach. data were collected on 8 patients performing nasopharyngeal examination. then reformat using 4 kernel variations. Respondens as many as 3 people radiologist specialist. Data analysis was done by using kappa test and friedman different test with spss version 16.
Result of this research indicate that there are difference of image from four variations kernel with p-value < 0.05. from the average value based on analysis and statistical test, the best image is to use H42S kernel.


Informasi Detail

Pernyataan Tanggungjawab
ARDI SOESILO WIBOWO, ST., DAN SIGIT WIJOKONGKO, S.SI, S.ST, M.KES
Pengarang
HAFID PRIHUTOMO - Pengarang Utama
NIM
P1337430216108
Bahasa
Indonesia
Deskripsi Fisik
-
Dilihat sebanyak
435
Penerbit Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG : POLTEKKES KEMENKES SEMARANG.,
Edisi
Subjek
Klasifikasi
081

Lampiran Berkas

Citation
HAFID PRIHUTOMO. (2017).PERBEDAAN INFORMASI CITRA ANATOMI PADA PEMERIKSAAN MSCTNASOFARING POTONGAN AXIAL DENGAN MENGGUNAKAN VARIASI KERNEL().POLTEKKES KEMENKES SEMARANG:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG

HAFID PRIHUTOMO.PERBEDAAN INFORMASI CITRA ANATOMI PADA PEMERIKSAAN MSCTNASOFARING POTONGAN AXIAL DENGAN MENGGUNAKAN VARIASI KERNEL().POLTEKKES KEMENKES SEMARANG:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2017.Skripsi D IV

HAFID PRIHUTOMO.PERBEDAAN INFORMASI CITRA ANATOMI PADA PEMERIKSAAN MSCTNASOFARING POTONGAN AXIAL DENGAN MENGGUNAKAN VARIASI KERNEL().POLTEKKES KEMENKES SEMARANG:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2017.Skripsi D IV

HAFID PRIHUTOMO.PERBEDAAN INFORMASI CITRA ANATOMI PADA PEMERIKSAAN MSCTNASOFARING POTONGAN AXIAL DENGAN MENGGUNAKAN VARIASI KERNEL().POLTEKKES KEMENKES SEMARANG:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2017.Skripsi D IV