Image of ANALISA PERBEDAAN VARIASI PERSENTASE NILAI ASIR TERHADAP NOISE DAN INFORMASI CITRA ANATOMI PADA PENGGUNAAN MODALITAS CT SCAN GENERAL ELECTRIC OPTIMA 660  :

(STUDI EKSPERIMEN PADA PENCITRAAN CT-SCAN ABDOMEN)

Skripsi D IV

ANALISA PERBEDAAN VARIASI PERSENTASE NILAI ASIR TERHADAP NOISE DAN INFORMASI CITRA ANATOMI PADA PENGGUNAAN MODALITAS CT SCAN GENERAL ELECTRIC OPTIMA 660 : (STUDI EKSPERIMEN PADA PENCITRAAN CT-SCAN ABDOMEN)

XML

Background : Adaptive Statistical Iterative Reconstruction (ASIR) is one of the software used to significantly reduce noise and increase image quality. Many various adjusment of percentage ASIR value in Imaging CT Scan Abdomen. The purpose of this research is to know the difference of noise and information of anatomical image on percentage variation of value ASIR and that reveal optimal anatomical image information of CT scan Abdomen.
Methods :The research type was experiment, the data was taken from 10 samples of image reconstruction of abdominal CT scan by giving four percentage variation of ASIR value (0%, 40%, 60%, and 80%). Measurement of noise value was done by placing ROI size 105.61 mm2 at three points are superior hepar, inferior liver and mid aorta on axial slices. As for the assessment of anatomical image information with observation of the result of variation of ASIR value by two radiologists. Data analysis using One Way Anova test to know difference of noise value, Friedman test to know difference of information of anatomical image on ASIR value variation and Friedman Mean Rank test to find optimal image information between percentage variation of ASIR value with 95% confidence level.
Results : The results showed that there was significant difference image of abdominal CT scan on percentage variation of ASIR value with p value


Informasi Detail

Pernyataan Tanggungjawab
dr. Lydia Purna Sp.Rad; Rini Indrati
Pengarang
Arieyanti Biyono - Pengarang Utama
NIM
P1337430216093
Bahasa
Indonesia
Deskripsi Fisik
BAB I - BAB V 60 HALAMAN, 29 cm
Dilihat sebanyak
1211
Penerbit Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG : Semarang.,
Edisi
Subjek
Klasifikasi
081

Lampiran Berkas

Citation
Arieyanti Biyono. (2017).ANALISA PERBEDAAN VARIASI PERSENTASE NILAI ASIR TERHADAP NOISE DAN INFORMASI CITRA ANATOMI PADA PENGGUNAAN MODALITAS CT SCAN GENERAL ELECTRIC OPTIMA 660 : (STUDI EKSPERIMEN PADA PENCITRAAN CT-SCAN ABDOMEN)().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG

Arieyanti Biyono.ANALISA PERBEDAAN VARIASI PERSENTASE NILAI ASIR TERHADAP NOISE DAN INFORMASI CITRA ANATOMI PADA PENGGUNAAN MODALITAS CT SCAN GENERAL ELECTRIC OPTIMA 660 : (STUDI EKSPERIMEN PADA PENCITRAAN CT-SCAN ABDOMEN)().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2017.Skripsi D IV

Arieyanti Biyono.ANALISA PERBEDAAN VARIASI PERSENTASE NILAI ASIR TERHADAP NOISE DAN INFORMASI CITRA ANATOMI PADA PENGGUNAAN MODALITAS CT SCAN GENERAL ELECTRIC OPTIMA 660 : (STUDI EKSPERIMEN PADA PENCITRAAN CT-SCAN ABDOMEN)().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2017.Skripsi D IV

Arieyanti Biyono.ANALISA PERBEDAAN VARIASI PERSENTASE NILAI ASIR TERHADAP NOISE DAN INFORMASI CITRA ANATOMI PADA PENGGUNAAN MODALITAS CT SCAN GENERAL ELECTRIC OPTIMA 660 : (STUDI EKSPERIMEN PADA PENCITRAAN CT-SCAN ABDOMEN)().Semarang:Prodi DIV T. Radiodiagnostik dan Radioterapi Semarang POLTEKKES KEMENKES SEMARANG,2017.Skripsi D IV



Dirujuk oleh 0 dokumen