Image of ANALISA PENERAPAN FAKTOR EKSPOSI ATURAN 15 % TERHADAP KUALITAS RADIOGRAF DAN DOSIS RADIASI RADIOGRAFI PELVIS PADA PENCITRAAN DIRECT DIGITAL RADIOGRAPHY

Tugas Akhir DIII

ANALISA PENERAPAN FAKTOR EKSPOSI ATURAN 15 % TERHADAP KUALITAS RADIOGRAF DAN DOSIS RADIASI RADIOGRAFI PELVIS PADA PENCITRAAN DIRECT DIGITAL RADIOGRAPHY

XML

Research on radiograph quality is image resolution resulting from standard exposure factor on direct digital radiography imaging at RSUD Moewardi Surakarta Radiology installation of 85 kVp and 20 mAs, and modification of exposure factor using 15% rule is 98 kVp 10 mAs and 72 kVp 40 MAs. This study aims to determine the radiograph quality of image resolution and radiation dose received by the detector with the exposure factor of 15% rule.
The type of this research is quantitative with experimental approach done on direct digital radiography imaging in radiology installation of RSUD Moewardi Surakarta using pelvic phantom and bar tool pattern and exposure three times using exposure factor standard 85 kVp 20 mAs and modification 15% rule that is 98 kVp 10 mAs and 72 kVp 40 mAs. After that the pelvic radiograph of the pelvis AP in read by respondents with three respondents and the data in can be made graph and then in descriptive analysis. Look for the dose of radiation received by the detector by calculating the intensity.
Based on the result of the reading of the bar pattern by the respondent of image resolution on the radiograph of the modified exposure factor modification factor 15% is 98 kVp 10 mAs obtained values of 1.6 and 2.13, and 72 kVp 40 mAs obtained 2.0 and 1.8. Which means Ha is rejected, so there is a 15% rule relationship to radiographic quality. At the dose of radiation received by the detector at 98 kVp 10 mAs obtained results 4.8 x 10-11 μGy and at 72 kVp 40 mAs obtained results 10.4 x 10-11 μGy.
Keyword : 15% rule, pelvis, image resolution, dose of radiation, direct digital radiography


Informasi Detail

Pernyataan Tanggungjawab
Dwi Rochmayanti, S.ST., M.Eng
Pengarang
WENING ANDAYANI - Pengarang Utama
NIM
P1337430114003
Bahasa
Indonesia
Deskripsi Fisik
xiv + 55 p.; Bibl.; Ilus.; 21 cm x 29,7 cm.
Dilihat sebanyak
781
Penerbit Poltekkes Kemenkes Semarang : Semarang.,
Edisi
Subjek
Klasifikasi
NONE

Lampiran Berkas

Citation
WENING ANDAYANI. (2017).ANALISA PENERAPAN FAKTOR EKSPOSI ATURAN 15 % TERHADAP KUALITAS RADIOGRAF DAN DOSIS RADIASI RADIOGRAFI PELVIS PADA PENCITRAAN DIRECT DIGITAL RADIOGRAPHY().Semarang:Poltekkes Kemenkes Semarang

WENING ANDAYANI.ANALISA PENERAPAN FAKTOR EKSPOSI ATURAN 15 % TERHADAP KUALITAS RADIOGRAF DAN DOSIS RADIASI RADIOGRAFI PELVIS PADA PENCITRAAN DIRECT DIGITAL RADIOGRAPHY().Semarang:Poltekkes Kemenkes Semarang,2017.Tugas Akhir DIII

WENING ANDAYANI.ANALISA PENERAPAN FAKTOR EKSPOSI ATURAN 15 % TERHADAP KUALITAS RADIOGRAF DAN DOSIS RADIASI RADIOGRAFI PELVIS PADA PENCITRAAN DIRECT DIGITAL RADIOGRAPHY().Semarang:Poltekkes Kemenkes Semarang,2017.Tugas Akhir DIII

WENING ANDAYANI.ANALISA PENERAPAN FAKTOR EKSPOSI ATURAN 15 % TERHADAP KUALITAS RADIOGRAF DAN DOSIS RADIASI RADIOGRAFI PELVIS PADA PENCITRAAN DIRECT DIGITAL RADIOGRAPHY().Semarang:Poltekkes Kemenkes Semarang,2017.Tugas Akhir DIII



Dirujuk oleh 0 dokumen